Lab Products
ProductsLabPhotometry > The Portable Photometers
You are looking for a dealer?
Please insert your query:

A useful note for field work:

For taking along all necessary utensils, such as test kits and spray bottle with distilled water as well as a disposal container, you can also pick a tool box from any from any building center to perfectly suit your needs.

Software

PC-Software LSData:
Convenient data management for the pHotoFlex®/Turb® 430 series
more ...

Please note:

As floating and moving particles are measured in turbidity, slight measurement deviations are possible. In order to achieve results that are as representative as possible, attention should be paid to the following:
samples should be measured immediately, as particles otherwise settle.
constant lamp operating temperature.
condensation on samples should be avoided.
the position of the standards should be marked to exclude the influence of glass inhomogenities.

further Products

Spectral and special analysis in industry, education and science and analysis of routine measurements with standard parameters in wastewater and drinking water, as well as environmental analysis and in-the-field use: Spectrophotometers of the photoLab® 6000 series
more ...

further Products

Thermoreactors for COD and all other thermal digestion processes
more …

pHotoFlex®: The Portable Photometers

The pHotoFlex® series offers the most robust optics, combining precision with low power consumption achieved through optical filters together with the LEDs. The instruments are equipped with 6 wavelengths. Additionally, the pH measuring and the optional turbidity measuring (IR range) are integrated, making these instruments the perfect partners for all measurements in the field: in a wastewater plant for wastewater and reference measurements, in drinking water analysis at a wellhead or in a cistern, and for monitoring bodies of water. They are versatile, low current and offer many extra features.

pHotoFlex® – Portable Photometer with pH

  • More than 160 methods available
  • Integrated pH measurement
  • Color measurements

The portable photometer pHotoFlex® demonstrates its capability with complex tasks in environmental and process monitoring at a variety of sites.
The integrated pH function allows measurements of pH 0 ... 16 with automatic buffer recognition (TEC/NIST). Temperature compensation is automatic within the permitted range of – 5 ... 100 °C (23 … 212 °F). WTW‘s MultiCal®-routine allows the automatic calibration with up to 3 calibration points. WTW offers a large selection of pH sensors as optional accessories: For field use, the maintenance-free SenTix® 41 is recommended, whereas for precision measurements in the laboratory, the SenTix® 81 glass electrode could be used. The electrodes are described in detail in the pH measuring chapter.

Light source   LED
Wavelengths nm   436, 517, 557, 594, 610, 690
User-defined methods   100
Methods/software update   via Internet
Data storage   1000 data sets
pH   0-16
Accuracy Photometry

pH
<2 nm wavelength accuracy,
0.005 abs. reproducibility
±0.01 pH

pHotoFlex® Turb – Total Capability

  • like pHotoFlex®, additionally:
  • Turbidity measurement according to DIN 27027/ ISO 7027
  • 0-1100 NTU/FNU
  • Calibration kit (0.02-10-1000 NTU)

The pHotoFlex® Turb is analogous to the pHotoFlex®, but includes an infrared (IR) light source for nephelometric turbidity measurement (90°), according to the requirements of DIN 27027/ISO 7027. Its precision is comparable to laboratory instruments for turbidity measurement.
The calibration with the supplied AMCO Clear® standards and measured data can be documented and output via RS232.
The AMCO Clear® standards enable highest precision for the sensitive testing of drinking water.

Light source   LED
Wavelengths nm   436, 517, 557, 594, 610, 690 + 860
User-defined methods   100
Methods/software update   via Internet
Data storage   1000 data sets
pH   0-16
Turbidity   0-1100 NTU/FNU
Accuracy Photometry


pH / Turbidity
< 2 nm wavelength accuracy,
0.005 abs. reproducibility
±0.01 pH
0.01 NTU/FNU or ±2% of the measured value

pHotoFlex® series in a convenient field case

  • The “in-field laboratory”
  • Integrated tray
  • Convenient

pHotoFlex® series in a convenient field case: A small lab for in-field use. The integrated tray features places for the instrument, cuvettes, measuring beaker and a stand for the pH electrode, making it practical for transport. Complete sets with:

  • pH electrode SenTix® 41 for all pHotoFlex® models
  • 1 variable pipette with 5 ml volume for all pHotoFlex® models
  • Calibration standards for pHotoFlex® Turb and Turb® 430 IR/T
  • LSdata for convenient data management and definition of user-defined methods.
  • Many useful accessories: empty cuvettes, buffer solutions with pH 4.01 and 7.00, PC cable AK Labor 540 B, stand for the pH electrode, cleaning tissues, screwdriver for battery change
  • Space for other accessories

LabStation